Search results for "Double lumen tube"

showing 1 items of 1 documents

Spiral-CT und 3 D-Rekonstruktion des Tracheobronchialbaumes zur Auswahl geeigneter Doppellumentuben vor Einzellungenbeatmung

1997

Purpose To assess the potential of 3-D spiral CT in the selection of adequate double-lumen tubes (DLT) for one-lung-ventilation. Materials and methods 3-D spiral CT of eight different DLTs was performed with standardized acquisition parameters: section thickness 4 mm, table speed 6 mm/s, increment 3 mm and standardised thresholds, window settings and magnification zoom. The accuracy was confirmed by measuring 3-D objects on screen and original DLTs by vernier caliper. 3-D spiral CT was performed in 20 patients preoperatively with slice thickness 4-8 mm, pitch 1-1.5, increment 4-8 mm. To select the adequate DLT size the 3-D reconstructions of the patient's tracheobronchial tree and of the DL…

medicine.medical_specialtySpiral CT ScansMaterials sciencebusiness.industrySlice thicknessMagnificationSurgerymedicineSuperimpositionCalipersRadiology Nuclear Medicine and imagingTomographyNuclear medicinebusinessSpiral ctDouble lumen tubeRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren
researchProduct